Aprašymas
Loop-powered temperature transducer for Pt 100 temperature sensors, configured via DIP switches, with spring-cage connection, pre-configured (please observe the order key). Replacement part: 2810308 MINI MCR-SL-PT100-LP-NC.
Product type | Temperature transmitter |
Insulation characteristics | |
Overvoltage category | II |
Pollution degree | 2 |
Rated insulation voltage | 30 V AC |
50 V DC | |
Electrical isolation | Basic insulation in accordance with EN 61010 |
Linearity error | < 0.05 % (for full measuring range) |
Test voltage, input/output/supply | 1.5 kV AC (50 Hz, 60 s) |
Step response (0–99%) | < 200 ms |
Maximum temperature coefficient | < 0.02 %/K |
Transmission error in the set measuring range | ((90 K / set measuring range [K]) + 0.05)% |
Transmission error in the full measuring range | ≤ 0,25 % |
Supply | |
Designation | Loop-powered |
Supply voltage range | 12 V DC … 30 V DC |
Max. current consumption | < 4.5 mA (without signal current) |
Power consumption | < 150 mW (without signal current) |
Signal | |
Number of inputs | 1 |
Input signal | Temperature |
Measurement | |
Configurable/programmable | Yes, preconfigured |
Sensor types (RTD) that can be used | Pt 100 (IEC 60751/EN 60751) |
Temperature measuring range | min. 50 K |
Sensor type: | -150 °C … 300 °C (configurable) |
Sensor input current | 1 mA (constant) |
Max. permissible overall conductor resistance | 10 Ω (Per cable) |
Connection technology | 2-, 3-, 4-conductor |
Error detection limit (short-circuit) | < 30 Ω |
Error detection limit (underrange) | 30 Ω ≤ (start span – 2.5% span) |
Error detection limit (overrange) | (end span + 2.5% span) ≤ approx. 254 Ω |
Error detection limit (wire break) | > approx. 254 Ω |
Signal | |
Number of outputs | 1 |
Configurable/programmable | Yes, preconfigured |
Current output signal | 4 mA … 20 mA |
20 mA … 4 mA | |
Max. current output signal | 23 mA (output limit) |
Load/output load current output | (Usupply – 12 V) / 22 mA |
Ripple | < 20 mVPP (at 500 Ω) |
Connection method | Spring-cage connection |
Stripping length | 8 mm |
Conductor cross section solid | 0.2 mm² … 2.5 mm² |
Conductor cross section flexible | 0.2 mm² … 2.5 mm² |
Conductor cross section AWG | 24 … 12 |
Dimensional drawing | |
Width | 6.2 mm |
Height | 93.1 mm |
Depth | 102.5 mm |
Fire protection for rail vehicles (DIN EN 45545-2) R22 | HL 1 – HL 2 |
Fire protection for rail vehicles (DIN EN 45545-2) R23 | HL 1 – HL 2 |
Fire protection for rail vehicles (DIN EN 45545-2) R24 | HL 1 – HL 2 |
Housing material | PBT |
Ambient conditions | |
Degree of protection | IP20 |
Ambient temperature (operation) | -20 °C … 65 °C |
Ambient temperature (storage/transport) | -40 °C … 85 °C |
Altitude | ≤ 2000 m |
Permissible humidity (operation) | 5 % … 95 % (non-condensing) |
CE | |
Certificate | CE-compliant |
ATEX | |
Identification | II 3 G Ex nA IIC T4 Gc X |
UL, USA / Canada | |
Identification | UL 508 Recognized |
Class I, Div. 2, Groups A, B, C, D T4 |
Electromagnetic compatibility | Conformance with EMC directive |
Noise emission | EN 61000-6-4 |
Noise immunity | EN 61000-6-2 |
Note | When being exposed to interference, there may be minimal deviations. |
Electrostatic discharge | |
Standards/regulations | EN 61000-4-2 |
Electrostatic discharge | |
Comments | Safety measures must be taken to prevent electrostatic discharge. |
Electromagnetic HF field | |
Designation | Electromagnetic RF field |
Standards/regulations | EN 61000-4-3 |
Typical deviation from the measuring range final value | 5 % |
Fast transients (burst) | |
Designation | Fast transients (burst) |
Standards/regulations | EN 61000-4-4 |
Typical deviation from the measuring range final value | 5 % |
Surge current load (surge) | |
Standards/regulations | EN 61000-4-5 |
Surge current load (surge) | |
Comments | Criterion B |
Conducted interference | |
Designation | Conducted interferences |
Standards/regulations | EN 61000-4-6 |
Typical deviation from the measuring range final value | 5 % |
Electrical isolation | Basic insulation in accordance with EN 61010 |
Mounting type | DIN rail mounting |
Mounting position | any |
Item number | 2810382 |
Packing unit | 1 pc |
Minimum order quantity | 1 pc |
Note | Made to Order (non-returnable) |
Sales Key | CK1 |
Product Key | CK1222 |
Catalog Page | Page 106 (C-7-2015) |
GTIN | 4046356151993 |
Weight per Piece (including packing) | 87.6 g |
Weight per Piece (excluding packing) | 87.6 g |
Customs tariff number | 85437090 |
Country of origin | DE |
ECLASS
ECLASS-9.0 | 27210129 |
ECLASS-10.0.1 | 27210129 |
ECLASS-11.0 | 27210129 |
ETIM
ETIM 7.0 | EC002919 |
2-, 3-, 4-conductor Pt 100 sensors|Highly-compact loop-powered temperature transducer for electrical isolation, conversion, amplification, and filtering of Pt 100 signals to create standard signals|Does not require additional auxiliary voltage|Error indication via diagnostic LED and analog signal|Input signals can be configured via DIP switches|2-way isolation|Temperature measuring range of -150°C to +300°C|Supplied by an output loop